| Anti Fouling |
UltraClean™ ultrasound cleaning system |
| Certifications |
cETLus, CE |
| Depth / Pressure Rating / Limit |
1 bar (14.5 psi) |
| Equipment used with |
282/284 and 2020 XT |
| Integrated Lightning Protection |
Yes |
| Languages |
English/ German |
| Measurement Method / Channels |
The probe measures the absorption at two wavelengths: 254 nm and 550 nm to compensate the SAC measurement for interferences from turbidity. |
| Measurement Range |
UVT-254: IQ UVT-254 701 (high range SAC values) 0.0 to 100.0 %; IQ UVT-254 705 (low range SAC values) 0.0 to 100.0 % SAC 254 soluble: IQ UVT-254 701 (high range SAC values) 0 to 3,000 1/m; IQ UVT-254 705 (low range SAC values) 0.0 to 600.0 1/m SAC-254 total: IQ UVT-254 701 (high range SAC values)0.0 to 3,000 1/m; IQ UVT-254 705 (low range SAC values) 0.0 to 600.0 1/m COD dissolved (correlated) 0.0 to 12,500 mg/L or 0.0 to 800.0 mg/L COD total (correlated) 0.0 to 20,000 mg/L or 0.0 to 800.0 mg/L TOC (correlated) 0.0 to 20,000 mg/L or 0.0 to 500.0 mg/L BOD (correlated) 0.0 to 8,000 mg/L or 0.0 to 500.0 mg/L DOC (correlated) 0.0 to 12,500 mg/L or 0.0 to 500.0 mg/L |
| Operating pH Range |
4 to 12 |
| Operating Temperature |
32 to 113 °F (0 to 45 °C) |
| Parameters Measured |
UVT-254, SAC (total and soluble) Correlated Parameters: COD, TOC, BOD, DOC *Only one parameter can be displayed at a time.
|
| Sampling |
Yes |
| Smart Sensors / Ports |
Yes |
| Storage Temperature |
14 to 122 °F (-10 to 50 °C) |
| Unit of Measure |
%, mg/L, 1/m |
| Warranty |
2 years |
| Waterproof |
Sensor with SACIQ cable connected: IP68 |