| Accuracy |
MR1: ± 3 %, ± 0.02 mg/l
MR2: ± 3 %, ± 0.10 mg/l |
| Calibration |
Automatic 1- and 2-point calibrations |
| Cleaning |
Automatic cleaning with cleaning solutions |
| Connectivity / Communications |
interfaces via IQ Sensor Net-components, relays, Modbus TCP, analog outputs |
| Dimensions |
30.5 x 26.6 x 17.5 in. |
| Dissolved Solids Limit |
< 6 g/L (before filtration) |
| Equipment used with |
IQ SensorNet 2020 3G Controller, 282/284 Controller |
| Languages |
English, German |
| Measurement Method |
Indophenol method |
| Measurement Range |
MR1: 0.02 to 4.00 mg/l NH4-N
MR2: 0.10 to 20.00 mg/l NH4-N |
| Operating Temperature |
-4 to 104 °F (-20 to +40) °C |
| Parameters Measured |
Ammonium |
| Power |
120 VAC / 240 VAC, 50/60 Hz |
| Reagent Use |
15 µL per measurement |
| Resolution |
MR1: 0.01 mg/l NH4-N
MR2: 0.05 mg/l NH4-N |
| Sample Streams/Channels |
1- and 2-channel versions available |
| Warranty |
2 Years |
| Weight |
81.6 lbs (37 kg) (without liquids) |